Cross-Calibration of Bone Mineral Density and Body Composition between GE Lunar Prodigy and Hologic Horizon W Dual Energy X-Ray Absorptiometry Systems

Journal of Bone Metabolism 2026;33(1):84-93.
Published online: February 28, 2026

1Department of Radiology, Chung-Ang University Hospital, Chung-Ang University College of Medicine, Seoul, Korea

2Department of Radiology, Chung-Ang University Gwangmyeong Hospital, Chung-Ang University College of Medicine, Gwangmyeong, Korea

3Department of Orthopedic Surgery, Inha University Hospital, Inha University School of Medicine, Incheon, Korea

4Department of Orthopedic Surgery, Busan Bumin Hospital, Busan, Korea

5Department of Orthopedic Surgery, Seoul Bumin Hospital, Seoul, Korea

Corresponding authors: Yong-Chan Ha, Department of Orthopedic Surgery, Seoul Bumin Hospital, 389 Gonghang-daero, Gangseo-gu, Seoul 07590, Korea, Tel: +82-2-2620-0058, Fax: +82-2-2620-0100, E-mail: hayongch@naver.com
Sang-Wook Lee, Department of Orthopedic Surgery, Busan Bumin Hospital, 59 Mandeok-daero, Buk-gu, Busan 46555, Korea, Tel: +82-51-330-3000, Fax: +82-51-330-3075, E-mail: osdrwook@gmail.com

Yong-Chan Ha and Sang-Wook Lee contributed equally to this work and should be considered co-corresponding authors.

• Received: January 14, 2026   • Revised: February 4, 2026   • Accepted: February 5, 2026

Copyright © 2026 The Korean Society for Bone and Mineral Research

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (https://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

  • 3,631 Views
  • 88 Download
  • 3 Crossref
  • 2 Scopus
prev next
  • Background
    Dual energy X-ray absorptiometry (DXA) measurements vary among instruments from different manufacturers. This study aimed to assess the correlation between bone mineral density (BMD) and body composition measured by the GE Lunar Prodigy and Hologic Horizon W DXA systems and to develop conversion equations.
  • Methods
    A total of 120 healthy volunteers (59 males, 61 females) aged 20 to 70 years were scanned on both DXA systems on the same day. BMD was measured at the lumbar spine (L1–L4) and bilateral femurs. Body composition parameters were measured for the whole body and regional sites. Bland-Altman analysis, Pearson correlation, and concordance correlation coefficient (CCC) were used to assess agreement. Linear regression was performed to derive conversion equations.
  • Results
    The Hologic system measured BMD values 14% to 20% lower at the lumbar spine and 8% to 17% lower at femoral sites compared to the GE Lunar system (all P<0.0001). Lean mass values were 4% to 10% lower on the Hologic system. Very strong correlations were observed for both BMD (r=0.944–0.980) and lean mass (r=0.963–0.984). Cross-calibration equations were developed for all measurement sites.
  • Conclusions
    Despite significant systematic differences, there was very high correlation between the two DXA systems. The conversion equations can facilitate comparison of measurements between these systems in clinical practice and research.
Osteoporosis is a systemic skeletal disorder characterized by decreased bone mass and microarchitectural deterioration of bone tissue, leading to enhanced bone fragility and an increased risk of fractures.[1]
Dual energy X-ray absorptiometry (DXA) has been established as the gold standard technique for measuring bone mineral density (BMD) since its introduction in the late 1980s.[2] DXA offers several advantages including high precision, low radiation exposure, rapid acquisition time, and the ability to measure BMD at clinically important skeletal sites.[3]
Beyond bone densitometry, DXA has emerged as a valuable tool for body composition assessment. The technique can accurately quantify lean mass, fat mass, and bone mineral content at both whole-body and regional levels.[4] This capability has become increasingly important with the recognition of sarcopenia—the age-related loss of skeletal muscle mass and function—as a significant contributor to adverse health outcomes in older adults.[5] The European Working Group on Sarcopenia in Older People (EWGSOP) has endorsed DXA as the preferred method for assessing muscle mass in clinical practice and research.[6]
The two dominant manufacturers of DXA systems are GE Healthcare (Madison, WI, USA) and Hologic Inc. (Bedford, MA, USA), which together account for the majority of DXA installations worldwide.[7] Although both manufacturers employ similar dual energy X-ray technology, significant differences exist in their calibration standards, proprietary algorithms for calculating BMD and body composition, edge detection methods, and reference databases.[8] Previous studies have demonstrated that Hologic systems typically report BMD values approximately 10% to 15% lower than GE Lunar systems at the spine and hip.[9]
These inter-manufacturer differences pose significant challenges in clinical practice and research. Early efforts to address inter-manufacturer differences led to the development of standardized BMD equations by the International Committee for Standards in Bone Measurement (ICSBM) in 1994.[10] However, these equations were derived from first-generation pencil-beam DXA systems, and subsequent studies have shown that they may not adequately eliminate differences between modern fan-beam systems. [10,11] Additionally, standardization equations have not been widely developed for body composition measurements, particularly for newer DXA models.[12]
The GE Lunar Prodigy and Hologic Horizon W represent current-generation DXA systems widely used in clinical practice. While several cross-calibration studies have examined older models from these manufacturers, data comparing these specific systems remain limited, particularly in Asian populations.[13] Given the differences in body composition and bone geometry between Asian and Western populations, population-specific conversion equations may be necessary for optimal accuracy.[14]
Therefore, the aims of this study were: (1) to investigate the correlation and agreement between BMD and body composition measurements obtained using the GE Lunar Prodigy and Hologic Horizon W DXA systems in healthy Korean adults; and (2) to develop cross-calibration equations that can facilitate the comparison and conversion of measurements between these two systems.
1. Study population
We conducted a prospective cross-calibration study at Chung-Ang University Hospital between August 2021 and October 2021. A total of 120 healthy volunteers (59 males and 61 females) who provided written informed consent were enrolled. The inclusion criteria were: (1) age between 20 and 70 years; (2) body mass index (BMI) between 18.5 and 35 kg/m2; and (3) ability to provide written informed consent. The exclusion criteria were: (1) underlying skeletal pathology such as scoliosis, osteoarthritis, diffuse idiopathic skeletal hyperostosis, or other clinical spinal deformities that could result in inaccurate BMD assessment; (2) previous history of adverse reaction after DXA scan or radiography; (3) pregnant or lactating women; (4) persons with artificial or implanted cardiac pacemakers; and (5) patients with mental illnesses such as severe depression that could affect their ability to cooperate during the examination. This study was approved by the Institutional Review Board (IRB) of Chung-Ang University Hospital (IRB No. 2021-023), and all participants provided written informed consent before enrollment.
2. DXA measurements
All participants underwent DXA scanning on both the GE Lunar Prodigy DXA system (GE Lunar Prodigy Advance; GE Medical Systems) and the Hologic Horizon W DXA system (Horizon W; Hologic Inc.) on the same day. The order of scanning on each system was randomized to minimize any potential systematic bias. All scans were performed according to each manufacturer’s standard positioning protocols.
For all scans, participants were positioned supine with lateral pronated hands by their side and both feet in 15 to 25 degrees of internal rotation. BMD measurements were obtained at the lumbar spine (L1–L4 vertebrae), bilateral femoral neck, trochanter, and total femur. Whole-body composition analysis was performed to obtain lean mass and fat mass values for bilateral upper extremities, bilateral lower extremities, trunk, and whole body. All measurements were automatically analyzed using each manufacturer’s standard software: GE Lunar enCORE version 17.0 for the GE system and Hologic Apex version 5.6 for the Hologic system. Daily quality assurance calibrations were performed on both systems according to manufacturer recommendations throughout the study period.
3. Statistical analysis
Descriptive statistics were presented as mean±standard deviation (SD) for continuous variables. Paired t-tests were used to compare measurements between the two DXA systems. The agreement between the two systems was assessed using multiple methods. First, Bland-Altman analysis was performed by plotting the differences between the two measurements against their mean values, with the calculation of bias (mean difference) and 95% limits of agreement (±1.96 SD).[15] Second, Pearson correlation coefficients were calculated to assess the linear relationship between measurements from the two systems, with correlations interpreted as weak (0.2–0.4), moderate (0.4–0.6), strong (0.6–0.8), or very strong (0.8–1.0).[16] Third, Lin’s concordance correlation coefficient (CCC) [17] was calculated to evaluate both the precision and accuracy of agreement, with CCC values graded as poor (<0.90), moderate (0.90–0.95), substantial (0.95–0.99), or almost perfect (>0.99).
Linear regression analysis was performed to derive conversion equations between the two systems, with measurements from one system as the independent variable and measurements from the other system as the dependent variable. Separate equations were developed for converting GE Lunar values to Hologic values and vice versa. All statistical analyses were performed using SPSS software version 26.0 (IBM Corp., Armonk, NY, USA). Statistical significance was set at P value less than 0.05.
1. Participant characteristics
The demographic characteristics of the 120 study participants are summarized in Table 1. The mean age was 48.6± 14.4 years, with a range from 20 to 70 years. The mean weight, height, and BMI were 65.43±12.36 kg, 165.45± 8.88 cm, and 23.78±3.28 kg/m2, respectively. Male participants (N=59) had a mean age of 47.6±13.9 years, weight of 72.70±11.75 kg, height of 172.16±5.30 cm, and BMI of 24.48±3.51 kg/m2. Female participants (N=61) had a mean age of 49.4±14.9 years, weight of 58.40±8.22 kg, height of 158.95±6.46 cm, and BMI of 23.11±2.90 kg/m2.
2. Comparison of BMD measurements
The BMD measurements from both DXA systems are presented in Table 2. All BMD values measured by the Hologic system were significantly lower than those measured by the GE Lunar system (all P<0.0001). At the lumbar spine, the mean BMD values ranged from 0.91±0.14 g/cm2 (L1) to 1.01±0.17 g/cm2 (L3) on the Hologic system, compared with 1.06±0.17 g/cm2 (L1) to 1.22±0.20 g/cm2 (L4) on the GE Lunar system. The mean difference for total L1–L4 BMD was 0.19 g/cm2 (16.4%), with the Hologic system measuring lower.
At the femoral sites, the Hologic BMD values ranged from 0.65±0.09 g/cm2 (left trochanter) to 0.89±0.12 g/cm2 (right total femur), while the GE Lunar values ranged from 0.78±0.13 g/cm2 (bilateral trochanter) to 0.97±0.14 g/cm2 (total femur). The mean differences between the two systems at the femoral sites ranged from 0.078 g/cm2 (8.0%) at the total femur to 0.156 g/cm2 (17.1%) at the femoral neck.
Bland-Altman analysis revealed consistent negative bias (Hologic − GE Lunar) across all BMD sites, with the mean differences ranging from −0.21 to −0.15 g/cm2 at the lumbar spine and from −0.16 to −0.08 g/cm2 at the femoral sites. The 95% limits of agreement for total L1–L4 BMD were −0.09 to −0.29 g/cm2 (Fig. 1, 2).
3. Comparison of body composition measurements
Lean mass measurements from both systems are shown in Table 3. Similar to BMD, the lean mass values measured by the Hologic system were systematically lower than those measured by the GE Lunar system at all sites (all P<0.0001). The mean whole-body lean mass was 41,055±8,681 g on the Hologic system compared with 44,143±9,279 g on the GE Lunar system, representing a mean difference of 3,088 g (7.0%). Regional differences were observed, with the greatest relative difference at the lower extremities (mean difference of 730 g or 10.4% for the right leg and 529 g or 7.7% for the left leg) and the smallest at the trunk (mean difference of 805 g or 3.8%).
Bland-Altman analysis showed that the magnitude of the difference between the two systems increased with increasing mean lean mass values, suggesting proportional bias. This pattern was consistent across all body regions measured.
4. Correlation and agreement analysis
The correlation and concordance statistics are presented in Table 4. Very strong Pearson correlations (r=0.944–0.984) were observed between the two systems for all BMD and lean mass measurements. For BMD, the correlations ranged from 0.944 (left femur trochanter) to 0.980 (L3 vertebra and right total femur). For lean mass, the correlations ranged from 0.963 (left lower extremity) to 0.984 (whole body).
Despite the strong correlations, the CCC showed greater variability, reflecting the systematic bias between the two systems. For BMD measurements, CCC values ranged from 0.505 (left femur trochanter) to 0.829 (right total femur), indicating poor agreement. The CCC was generally lower at the femoral neck and trochanter sites compared with the total femur and spine. For lean mass measurements, CCC values ranged from 0.854 (right upper extremity) to 0.946 (trunk), indicating poor to moderate agreement. The agreement was better for larger body regions (trunk and whole body) compared with smaller regions (extremities).
5. Cross-calibration equations
The conversion equations derived from linear regression analysis are presented in Table 5. For converting GE Lunar BMD values to Hologic values, the equations followed the general form: Hologic=(slope×GE Lunar)+intercept. The slopes ranged from 0.63 to 0.86, reflecting the proportionally lower values obtained with the Hologic system. For example, the equation for total L1–L4 BMD was: Hologic=(0.81×GE Lunar)+0.037.
For lean mass conversions, the slopes ranged from 0.84 to 0.93 for converting GE Lunar to Hologic values. The equation for whole-body lean mass was: Hologic= (0.92×GE Lunar)+398. Reverse equations (Hologic to GE Lunar) are also provided for clinical scenarios requiring conversion in the opposite direction.
This prospective cross-calibration study provides comprehensive data on the relationship between BMD and body composition measurements obtained using the GE Lunar Prodigy and Hologic Horizon W DXA systems in healthy Korean adults. Our findings confirm the existence of significant systematic differences between these two widely used DXA systems while demonstrating very strong correlations that support the feasibility of cross-calibration. The conversion equations derived in this study can facilitate the comparison and transfer of DXA data between these systems in both clinical practice and research settings.
The observed differences in BMD measurements between the GE Lunar and Hologic systems are consistent with previous cross-calibration studies, although the magnitude of differences varies depending on the specific models and skeletal sites examined. In our study, the Hologic system consistently measured BMD values approximately 14% to 20% lower than the GE Lunar system at the lumbar spine and 8% to 17% lower at the femoral sites. These findings align with the well-documented observation that Hologic systems typically report lower BMD values than GE Lunar systems.[9,10]
The clinical implications of these inter-manufacturer differences are substantial. Given that the World Health Organization diagnostic criteria for osteoporosis are based on T-score thresholds derived from BMD measurements, the use of different DXA systems could lead to different diagnostic classifications for the same patient.[18] A recent study by Analay et al. [19] demonstrated that among patients diagnosed with osteoporosis on a Hologic system, up to 51% would be reclassified as having only osteopenia when measured on a Lunar system. This discrepancy underscores the critical importance of using appropriate cross-calibration when patients are monitored on different systems or when DXA data from different sources are pooled for research purposes.
An important finding of our study is the differential magnitude of between-system differences across anatomical sites. The smallest relative differences were observed at the total femur (8.0%), while the largest differences were at the femoral neck and trochanter (14%–17%) and lumbar spine (16%–18%). This pattern has clinical relevance because different skeletal sites are often used for different purposes—the femoral neck is typically used for fracture risk assessment with fracture risk assessment tool, while the spine may be preferred for monitoring treatment response due to its higher metabolic activity.[20] The site-specific nature of the between-system differences highlights the importance of applying the appropriate conversion equation for each measurement site.
With the growing recognition of sarcopenia as a significant health concern in aging populations, the accurate assessment of lean mass has become increasingly important. [5,6] Our study demonstrates that the Hologic system consistently measures lean mass values 4% to 10% lower than the GE Lunar system across all body regions. These differences are clinically significant because they could affect the diagnosis of sarcopenia, which is based on cutoff values for appendicular lean mass.[6] Previous studies have reported similar magnitude of differences in lean mass measurements between different DXA systems.[12,13] Park et al. [13] recently published conversion equations between the same two DXA systems in Korean adults, and our results are generally consistent with their findings, providing cross-validation for the reliability of these conversion equations.
The observation that the between-system differences in lean mass increased with increasing mean values (proportional bias) has important implications. This pattern suggests that the relationship between the two systems is not simply a constant offset but involves both systematic and proportional components.
The very strong Pearson correlations (r>0.94 for all measurements) observed in our study indicate that the two systems are measuring the same underlying constructs with high fidelity. However, the CCC, which account for both precision and accuracy, were notably lower for several sites, reflecting the systematic bias between the systems. This discrepancy between correlation and concordance illustrates why simple correlation analysis is insufficient for method comparison studies.[15,16] The Bland-Altman approach and CCC provide complementary information about agreement that is essential for proper interpretation of cross-calibration data.
Several limitations of this study should be acknowledged. First, although our sample size of 120 participants is adequate for developing regression equations and is consistent with previous cross-calibration studies, larger samples would provide narrower confidence intervals for the conversion coefficients. Second, our study population consisted of healthy Korean adults, and the conversion equations may not be directly applicable to populations with different body compositions, ethnicities, or disease states. Third, while we included participants across a wide age range (20–70 years) and BMI range (18.5–35 kg/m2), extremely elderly individuals and those with morbid obesity were not represented. Fourth, we did not include T-score comparisons, which would require knowledge of the specific reference databases used by each manufacturer. Finally, our study examined only two specific DXA models, and the conversion equations may not be applicable to other models from the same manufacturers.
Despite these limitations, our study has several strengths. The prospective design with same-day scanning on both systems minimizes the confounding effects of biological variation that could occur if measurements were obtained on different days. The inclusion of both BMD and body composition measurements provides comprehensive cross-calibration data relevant to current clinical practice. The use of multiple statistical methods (paired t-tests, Bland-Altman analysis, Pearson correlation, and CCC) provides a thorough characterization of the agreement between the two systems.
This study demonstrates a very high correlation between BMD and body composition measurements obtained using the GE Lunar Prodigy and Hologic Horizon W DXA systems, despite the presence of significant systematic differences. The Hologic system consistently measures lower values for both BMD and lean mass compared to the GE Lunar system. The conversion equations derived in this study can facilitate the comparison and transfer of DXA data between these systems, which is valuable for longitudinal patient monitoring, multicenter clinical trials, and epidemiological research.

Funding

This research was supported by a fund (2021-11-007) by Research of Korea Disease Control and Prevention Agency.

Ethics approval and consent to participate

The study protocol conformed to the ethical guidelines of the 1975 Declaration of Helsinki and was approved by the Institutional Review Board of the Chung-Ang University Hospital (IRB No. 2021-023). Written informed consent was obtained from all participants.

Conflict of interest

No potential conflict of interest relevant to this article was reported.

Fig. 1
Bland-Altman plots showing the agreement between GE Lunar Prodigy and Hologic Horizon W dual energy X-ray absorptiometry systems for lumbar spine bone mineral density (BMD) measurements. The y-axis represents the difference (Hologic − GE Lunar) and the x-axis represents the mean of the two measurements. The solid line indicates the mean difference (bias), and the dashed lines indicate the 95% limits of agreement (±1.96 standard deviation [SD]). L, lumbar vertebra.
jbm-26-1014f1.jpg
Fig. 2
Bland-Altman plots showing the agreement between GE Lunar Prodigy and Hologic Horizon W dual energy X-ray absorptiometry systems for femoral bone mineral density (BMD) measurements. The y-axis represents the difference (Hologic − GE Lunar) and the x-axis represents the mean of the two measurements. The solid line indicates the mean difference (bias), and the dashed lines indicate the 95% limits of agreement (±1.96 standard deviation [SD]). U/E, upper extremity; L/E, lower extremity.
jbm-26-1014f2.jpg
jbm-26-1014f3.jpg
Table 1
Demographic characteristics of the study participants
Table 1
Variables Total (N=120) Male (N=59) Female (N=61)
Age (yr) 48.6±14.39 47.6±13.90 49.4±14.91
Weight (kg) 65.43±12.36 72.70±11.75 58.40±8.22
Height (m) 165.45±8.88 172.16±5.30 158.95±6.46
BMI (kg/m2) 23.78±3.28 24.48±3.51 23.11±2.90

The data is presented as mean±standard deviation. BMI, body mass index.

Table 2
Comparison of bone mineral density measurements between Hologic Horizon W and GE Lunar Prodigy dual energy X-ray absorptiometry systems
Table 2
BMD Hologic Horizon W GE Lunar Prodigy P-value


Mean±SD Range Mean±SD Range
L1 0.91±0.14 0.59–1.39 1.06±0.17 0.69–1.58 <0.0001a)

L2 0.97±0.15 0.61–1.53 1.14±0.18 0.72–1.74 <0.0001a)

L3 1.01±0.17 0.60–1.68 1.21±0.2 0.77–1.98 <0.0001a)

L4 1.00±0.17 0.64–1.58 1.22±0.2 0.78–1.91 <0.0001a)

Total L1–4 0.97±0.15 0.65–1.55 1.16±0.18 0.78–1.81 <0.0001a)

Right femur neck 0.76±0.12 0.51–1.17 0.91±0.13 0.63–1.35 <0.0001a)

Right femur trochanter 0.67±0.09 0.45–0.91 0.78±0.13 0.48–1.14 <0.0001a)

Right femur total 0.89±0.12 0.65–1.20 0.97±0.14 0.71–1.32 <0.0001a)

Left femur neck 0.76±0.12 0.50–1.08 0.91±0.13 0.58–1.29 <0.0001a)

Left femur trochanter 0.65±0.09 0.43–0.89 0.78±0.13 0.50–1.14 <0.0001a)

Left femur total 0.88±0.12 0.65–1.16 0.97±0.14 0.68–1.31 <0.0001a)

a)P<0.0001 by paired t-test.

BMD, bone mineral density; L, lumbar vertebra; SD, standard deviation.

Table 3
Comparison of lean mass measurements between Hologic Horizon W and GE Lunar Prodigy dual energy X-ray absorptiometry systems
Table 3
Lean mass Hologic Horizon W GE Lunar Prodigy P-value


Mean±SD Range Mean±SD Range
Right upper extremity 2,067.76±690.57 994–4,489.3 2,451.93±802.57 1,208–5,307 <0.0001a)

Left upper extremity 1,986.89±681.77 940–4,191.4 2,296.21±777.20 1,173–5,241 <0.0001a)

Right lower extremity 6,294.36±1,561.85 3,902–10,436.6 7,024.79±1,642.30 4,222–10,953 <0.0001a)

Left lower extremity 6,384.86±1,550.80 3,938–10,798.9 6,913.41±1,599.82 4,476–10,943 <0.0001a)

Trunk 20,438.58±4,043.51 13,196–31,429 21,243.10±4,246.83 13,559–31,472 <0.0001a)

Whole body 41,055.2±8,680.72 26,223–62,911.1 44,143.32±9,278.78 28,303–67,684 <0.0001a)

a)P<0.0001 by paired t-test.

SD, standard deviation.

Table 4
Correlation and concordance between GE Lunar Prodigy and Hologic Horizon W dual energy X-ray absorptiometry systems
Table 4
Pearson correlation (r) CCC 95% CI
BMD
 L1 0.959 0.652 0.585–0.711
 L2 0.964 0.636 0.567–0.695
 L3 0.98 0.594 0.526–0.654
 L4 0.967 0.559 0.488–0.622
 Total L1–4 0.975 0.584 0.515–0.645
 Right femur neck 0.956 0.537 0.466–0.602
 Right femur trochanter 0.958 0.597 0.527–0.659
 Right femur total 0.98 0.829 0.785–0.865
 Left femur neck 0.955 0.528 0.457–0.593
 Left femur trochanter 0.944 0.505 0.433–0.570
 Left femur total 0.977 0.756 0.701–0.802

Lean mass
 Right upper extremity 0.978 0.854 0.814–0.886
 Left upper extremity 0.972 0.884 0.849–0.911
 Right lower extremity 0.967 0.875 0.836–0.904
 Left lower extremity 0.963 0.911 0.880–0.934
 Trunk 0.965 0.946 0.925–0.961
 Whole body 0.984 0.927 0.904–0.945

BMD, bone mineral density; L, lumbar vertebra; CCC, concordance correlation coefficient; CI, confidence interval.

Table 5
Cross-calibration equations between GE Lunar Prodigy and Hologic Horizon W dual energy X-ray absorptiometry systems
Table 5
Conversion equation (GE Lunar Prodigy to Hologic Horizon W) Conversion equation (Hologic Horizon W to GE Lunar Prodigy)
BMD
 L1 Hol_L1=(0.82×GE_L1)+0.048 GE_L1=(1.13×Hol_L1)+0.030
 L2 Hol_L2=(0.81×GE_L2)+0.050 GE_L2=(1.15×Hol_L2)+0.022
 L3 Hol_L3=(0.83×GE_L3) − 0.004 GE_L3=(1.15×Hol_L3)+0.052
 L4 Hol_L4=(0.83×GE_L4) − 0.009 GE_L4=(1.13×Hol_L4)+0.090
 Total L1–4 Hol_L1–4=(0.81×GE_L1–4)+0.037 GE_L1–4=(1.18×Hol_L1–4)+0.014
 Right femur neck Hol_RFneck=(0.86×GE_RFneck) − 0.026 GE_RFneck=(1.18×Hol_RFneck)+0.107
 Right femur trochanter Hol_RFtro=(0.68×GE_RFtro)+0.137 GE_RFtro=(1.35×Hol_RFtro) − 0.121
 Right femur total Hol_RFTotal=(0.86×GE_RFTotal)+0.062 GE_RFTotal=(1.12×Hol_RFTotal) − 0.031
 Left femur neck Hol_LFneck=(0.85×GE_LFneck) − 0.019 GE_LFneck=(1.07×Hol_LFneck)+0.101
 Left femur trochanter Hol_LFtro=(0.63×GE_LFtro)+0.154 GE_LFtro=(1.41×Hol_LFtro) − 0.131
 Left femur total Hol_LFTotal=(0.82×GE_LFTotal)+0.081 GE_LFTotal=(1.16×Hol_LFTotal) − 0.050

Lean mass
 Right upper extremity Hol_RUE=(0.84×GE_RUE)+4 GE_RUE=(1.14×Hol_RUE)+102
 Left upper extremity Hol_LUE=(0.85×GE_LUE)+29 GE_LUE=(1.11×Hol_LUE)+95
 Right lower extremity Hol_RLE=(0.92×GE_RLE) −169 GE_RLE=(1.02×Hol_RLE)+622
 Left lower extremity Hol_LLE=(0.93×GE_LLE) − 70 GE_LLE=(0.99×Hol_LLE)+569
 Trunk Hol_Trunk=(0.92×GE_Trunk)+915 GE_Trunk=(1.01×Hol_Trunk)+522
 Whole body Hol_WB=(0.92×GE_WB)+398 GE_WB=(1.05×Hol_WB)+941

BMD, bone mineral density; L, lumbar vertebra; Hol, Hologic; GE, GE Lunar; RF, right femur; LF, left femur; RUE, right upper extremity; LUE, left upper extremity; RLE, right lower extremity; LLE, left lower extremity; WB, whole body.

Download Citation

Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

Format:

Include:

Cross-Calibration of Bone Mineral Density and Body Composition between GE Lunar Prodigy and Hologic Horizon W Dual Energy X-Ray Absorptiometry Systems
J Bone Metab. 2026;33(1):84-93.   Published online February 28, 2026
Download Citation

Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

Format:
Include:
Cross-Calibration of Bone Mineral Density and Body Composition between GE Lunar Prodigy and Hologic Horizon W Dual Energy X-Ray Absorptiometry Systems
J Bone Metab. 2026;33(1):84-93.   Published online February 28, 2026
Close

Figure

  • 0
  • 1
  • 2
Cross-Calibration of Bone Mineral Density and Body Composition between GE Lunar Prodigy and Hologic Horizon W Dual Energy X-Ray Absorptiometry Systems
Image Image Image
Fig. 1 Bland-Altman plots showing the agreement between GE Lunar Prodigy and Hologic Horizon W dual energy X-ray absorptiometry systems for lumbar spine bone mineral density (BMD) measurements. The y-axis represents the difference (Hologic − GE Lunar) and the x-axis represents the mean of the two measurements. The solid line indicates the mean difference (bias), and the dashed lines indicate the 95% limits of agreement (±1.96 standard deviation [SD]). L, lumbar vertebra.
Fig. 2 Bland-Altman plots showing the agreement between GE Lunar Prodigy and Hologic Horizon W dual energy X-ray absorptiometry systems for femoral bone mineral density (BMD) measurements. The y-axis represents the difference (Hologic − GE Lunar) and the x-axis represents the mean of the two measurements. The solid line indicates the mean difference (bias), and the dashed lines indicate the 95% limits of agreement (±1.96 standard deviation [SD]). U/E, upper extremity; L/E, lower extremity.
Graphical abstract
Cross-Calibration of Bone Mineral Density and Body Composition between GE Lunar Prodigy and Hologic Horizon W Dual Energy X-Ray Absorptiometry Systems

Demographic characteristics of the study participants

Variables Total (N=120) Male (N=59) Female (N=61)
Age (yr) 48.6±14.39 47.6±13.90 49.4±14.91
Weight (kg) 65.43±12.36 72.70±11.75 58.40±8.22
Height (m) 165.45±8.88 172.16±5.30 158.95±6.46
BMI (kg/m2) 23.78±3.28 24.48±3.51 23.11±2.90

The data is presented as mean±standard deviation. BMI, body mass index.

Comparison of bone mineral density measurements between Hologic Horizon W and GE Lunar Prodigy dual energy X-ray absorptiometry systems

BMD Hologic Horizon W GE Lunar Prodigy P-value


Mean±SD Range Mean±SD Range
L1 0.91±0.14 0.59–1.39 1.06±0.17 0.69–1.58 <0.0001a)

L2 0.97±0.15 0.61–1.53 1.14±0.18 0.72–1.74 <0.0001a)

L3 1.01±0.17 0.60–1.68 1.21±0.2 0.77–1.98 <0.0001a)

L4 1.00±0.17 0.64–1.58 1.22±0.2 0.78–1.91 <0.0001a)

Total L1–4 0.97±0.15 0.65–1.55 1.16±0.18 0.78–1.81 <0.0001a)

Right femur neck 0.76±0.12 0.51–1.17 0.91±0.13 0.63–1.35 <0.0001a)

Right femur trochanter 0.67±0.09 0.45–0.91 0.78±0.13 0.48–1.14 <0.0001a)

Right femur total 0.89±0.12 0.65–1.20 0.97±0.14 0.71–1.32 <0.0001a)

Left femur neck 0.76±0.12 0.50–1.08 0.91±0.13 0.58–1.29 <0.0001a)

Left femur trochanter 0.65±0.09 0.43–0.89 0.78±0.13 0.50–1.14 <0.0001a)

Left femur total 0.88±0.12 0.65–1.16 0.97±0.14 0.68–1.31 <0.0001a)

a)P<0.0001 by paired t-test.

BMD, bone mineral density; L, lumbar vertebra; SD, standard deviation.

Comparison of lean mass measurements between Hologic Horizon W and GE Lunar Prodigy dual energy X-ray absorptiometry systems

Lean mass Hologic Horizon W GE Lunar Prodigy P-value


Mean±SD Range Mean±SD Range
Right upper extremity 2,067.76±690.57 994–4,489.3 2,451.93±802.57 1,208–5,307 <0.0001a)

Left upper extremity 1,986.89±681.77 940–4,191.4 2,296.21±777.20 1,173–5,241 <0.0001a)

Right lower extremity 6,294.36±1,561.85 3,902–10,436.6 7,024.79±1,642.30 4,222–10,953 <0.0001a)

Left lower extremity 6,384.86±1,550.80 3,938–10,798.9 6,913.41±1,599.82 4,476–10,943 <0.0001a)

Trunk 20,438.58±4,043.51 13,196–31,429 21,243.10±4,246.83 13,559–31,472 <0.0001a)

Whole body 41,055.2±8,680.72 26,223–62,911.1 44,143.32±9,278.78 28,303–67,684 <0.0001a)

a)P<0.0001 by paired t-test.

SD, standard deviation.

Correlation and concordance between GE Lunar Prodigy and Hologic Horizon W dual energy X-ray absorptiometry systems

Pearson correlation (r) CCC 95% CI
BMD
 L1 0.959 0.652 0.585–0.711
 L2 0.964 0.636 0.567–0.695
 L3 0.98 0.594 0.526–0.654
 L4 0.967 0.559 0.488–0.622
 Total L1–4 0.975 0.584 0.515–0.645
 Right femur neck 0.956 0.537 0.466–0.602
 Right femur trochanter 0.958 0.597 0.527–0.659
 Right femur total 0.98 0.829 0.785–0.865
 Left femur neck 0.955 0.528 0.457–0.593
 Left femur trochanter 0.944 0.505 0.433–0.570
 Left femur total 0.977 0.756 0.701–0.802

Lean mass
 Right upper extremity 0.978 0.854 0.814–0.886
 Left upper extremity 0.972 0.884 0.849–0.911
 Right lower extremity 0.967 0.875 0.836–0.904
 Left lower extremity 0.963 0.911 0.880–0.934
 Trunk 0.965 0.946 0.925–0.961
 Whole body 0.984 0.927 0.904–0.945

BMD, bone mineral density; L, lumbar vertebra; CCC, concordance correlation coefficient; CI, confidence interval.

Cross-calibration equations between GE Lunar Prodigy and Hologic Horizon W dual energy X-ray absorptiometry systems

Conversion equation (GE Lunar Prodigy to Hologic Horizon W) Conversion equation (Hologic Horizon W to GE Lunar Prodigy)
BMD
 L1 Hol_L1=(0.82×GE_L1)+0.048 GE_L1=(1.13×Hol_L1)+0.030
 L2 Hol_L2=(0.81×GE_L2)+0.050 GE_L2=(1.15×Hol_L2)+0.022
 L3 Hol_L3=(0.83×GE_L3) − 0.004 GE_L3=(1.15×Hol_L3)+0.052
 L4 Hol_L4=(0.83×GE_L4) − 0.009 GE_L4=(1.13×Hol_L4)+0.090
 Total L1–4 Hol_L1–4=(0.81×GE_L1–4)+0.037 GE_L1–4=(1.18×Hol_L1–4)+0.014
 Right femur neck Hol_RFneck=(0.86×GE_RFneck) − 0.026 GE_RFneck=(1.18×Hol_RFneck)+0.107
 Right femur trochanter Hol_RFtro=(0.68×GE_RFtro)+0.137 GE_RFtro=(1.35×Hol_RFtro) − 0.121
 Right femur total Hol_RFTotal=(0.86×GE_RFTotal)+0.062 GE_RFTotal=(1.12×Hol_RFTotal) − 0.031
 Left femur neck Hol_LFneck=(0.85×GE_LFneck) − 0.019 GE_LFneck=(1.07×Hol_LFneck)+0.101
 Left femur trochanter Hol_LFtro=(0.63×GE_LFtro)+0.154 GE_LFtro=(1.41×Hol_LFtro) − 0.131
 Left femur total Hol_LFTotal=(0.82×GE_LFTotal)+0.081 GE_LFTotal=(1.16×Hol_LFTotal) − 0.050

Lean mass
 Right upper extremity Hol_RUE=(0.84×GE_RUE)+4 GE_RUE=(1.14×Hol_RUE)+102
 Left upper extremity Hol_LUE=(0.85×GE_LUE)+29 GE_LUE=(1.11×Hol_LUE)+95
 Right lower extremity Hol_RLE=(0.92×GE_RLE) −169 GE_RLE=(1.02×Hol_RLE)+622
 Left lower extremity Hol_LLE=(0.93×GE_LLE) − 70 GE_LLE=(0.99×Hol_LLE)+569
 Trunk Hol_Trunk=(0.92×GE_Trunk)+915 GE_Trunk=(1.01×Hol_Trunk)+522
 Whole body Hol_WB=(0.92×GE_WB)+398 GE_WB=(1.05×Hol_WB)+941

BMD, bone mineral density; L, lumbar vertebra; Hol, Hologic; GE, GE Lunar; RF, right femur; LF, left femur; RUE, right upper extremity; LUE, left upper extremity; RLE, right lower extremity; LLE, left lower extremity; WB, whole body.

Table 1 Demographic characteristics of the study participants

The data is presented as mean±standard deviation. BMI, body mass index.

Table 2 Comparison of bone mineral density measurements between Hologic Horizon W and GE Lunar Prodigy dual energy X-ray absorptiometry systems

P<0.0001 by paired t-test.

BMD, bone mineral density; L, lumbar vertebra; SD, standard deviation.

Table 3 Comparison of lean mass measurements between Hologic Horizon W and GE Lunar Prodigy dual energy X-ray absorptiometry systems

P<0.0001 by paired t-test.

SD, standard deviation.

Table 4 Correlation and concordance between GE Lunar Prodigy and Hologic Horizon W dual energy X-ray absorptiometry systems

BMD, bone mineral density; L, lumbar vertebra; CCC, concordance correlation coefficient; CI, confidence interval.

Table 5 Cross-calibration equations between GE Lunar Prodigy and Hologic Horizon W dual energy X-ray absorptiometry systems

BMD, bone mineral density; L, lumbar vertebra; Hol, Hologic; GE, GE Lunar; RF, right femur; LF, left femur; RUE, right upper extremity; LUE, left upper extremity; RLE, right lower extremity; LLE, left lower extremity; WB, whole body.