Although dual energy X-ray absorptiometry (DXA) is known to standard equipment for bone mineral density (BMD) measurements. Different results of BMD measurement using a number of different types of devices are difficult to use clinical practice. The purpose of this study was to evaluate discrepancy and standardizations of DXA devices from three manufactures using a European Spine Phantom (ESP).
We calculated the accuracy and precision of 36 DXA devices from three manufacturers (10 Hologic, 16 Lunar, and 10 Osteosys) using a ESP (semi-anthropomorphic). The ESP was measured 5 times on each equipment without repositioning. Accuracy was assessed by comparing BMD (g/cm2) values measured on each device with the actual value of the phantom. Precision was assessed by the coefficient of variation (CVsd).
Lunar devices were, on average, 22%, 8.3%, and 5% overestimation for low (L1) BMD values, medium (L2), and high (L3) BMD values. Hologic devices were, on average, 6% overestimation for L1 BMD, and 5% and 6.2% underestimation for L2 and L3 BMD values. Osteosys devices was, on average, 12.7% (0.063 g/cm2), 6.3% (0.062 g/cm2), and 5% (0.075 g/cm2) underestimation for L1, L2, and L3, respectively. The mean CVsd for L1-L3 BMD were 0.01%, 0.78%, and 2.46% for Lunar, Hologic, and Osteosys devices respectively.
The BMD comparison in this study demonstrates that BMD result of three different devices are significant different between three devices. Differences of BMD between three devices are necessary to BMD standardization.
Citations
In management of osteoporosis, several concerns here have been raised. The current issue included the utilization of dual energy X-ray absorptiometry (DXA) and fracture-risk assessment (FRAX), screening of vitamin D deficiency and secondary osteoporosis, and long-term use of bisphosphonate and calcium supplements. There was no study on physicians' attitude on these current issues in Korea. Therefore, we investigated the physicians' attitude on these issues by survey.
We administered a 30-item questionnaire to all members of Korean Society for Bone and Mineral Research by email survey form. One hundred participants answered the questionnaire. The questionnaire included the questions about the physicians' attitude to current issues and the barriers to osteoporosis treatment in Korea.
Most physicians used bone densitometry devices (99%) and, central DXA was the most accessible device (95%). Eighty-eight percent were aware of FRAX®, but among them, only 19.3% used it. The main reason for not using FRAX® was the lack of time in their proactive (76%). Screening for vitamin D status and secondary osteoporosis was performed by 59% and 52% of the respondents, respectively. The lack of awareness among patients and high costs of medication were perceived as the most important barriers to osteoporosis management in Korea.
This study provides physicians' perspective to the current issue for diagnostic and treatment of osteoporosis in Korea. To further improve osteoporosis management, educational programs for patients and doctors, and the improvement of reimbursement system should be considered in Korea.
Citations
The image quality management of bone mineral density (BMD) is the responsibility and duty of radio-technologists who carry out examinations. However, inaccurate conclusions due to the lack of understanding and ignorance regarding the methodology of image quality management can be a fatal error to patients. The accuracy and precision of BMD measurement must be maintained at the highest level so that actual biological changes can be detected with even slight changes in BMD. Accuracy and precision should be continuously preserved for image quality of machines. Those factors will contribute to ensure the reliability of BMD examination. The enforcement of proper quality control of radiologists performing BMD inspections which brings about the durability extensions of equipment and accurate results of calculations will help the assurance of reliable inspections.
Citations