Original Articles

Cross-Calibration of Bone Mineral Density and Body Composition between GE Lunar Prodigy and Hologic Horizon W Dual Energy X-Ray Absorptiometry Systems
Sujin Kim, Guen Young Lee, Jun-Il Yoo, Sang-Wook Lee, Yong-Chan Ha
J Bone Metab 2026;33(1):84-93.
Published online February 28, 2026
DOI: https://doi.org/10.11005/jbm.26.1014
Background
Dual energy X-ray absorptiometry (DXA) measurements vary among instruments from different manufacturers. This study aimed to assess the correlation between bone mineral density (BMD) and body composition measured by the GE Lunar Prodigy and Hologic Horizon W DXA systems and to develop conversion equations.
Methods
A total of 120 healthy volunteers (59 males, 61 females) aged 20 to 70 years were scanned on both DXA systems on the same day. BMD was measured at the lumbar spine (L1-L4) and bilateral femurs. Body composition parameters were measured for the whole body and regional sites. Bland-Altman analysis, Pearson correlation, and concordance correlation coefficient (CCC) were used to assess agreement. Linear regression was performed to derive conversion equations.
Results
The Hologic system measured BMD values 14% to 20% lower at the lumbar spine and 8% to 17% lower at femoral sites compared to the GE Lunar system (all P<0.0001). Lean mass values were 4% to 10% lower on the Hologic system. Very strong correlations were observed for both BMD (r=0.944-0.980) and lean mass (r=0.963-0.984). Cross-calibration equations were developed for all measurement sites.
Conclusions
Despite significant systematic differences, there was very high correlation between the two DXA systems. The conversion equations can facilitate comparison of measurements between these systems in clinical practice and research.

Citations

Citations to this article as recorded by  Crossref logo
  • 1. Comparison of adenoma and gland hyperplasia on bone mineral density in primary hyperparathyroidism
    Alexander Green, Kate Hardy, Oday Al-Dadah
    Journal of Clinical Densitometry.2026; 29(3): 101718.     CrossRef
  • 2. Bioimpedance: From Body Composition Assessment to Emerging Technologies in Clinical Nutrition
    Ruth Nekane Pita, Isabel María Fernández, Oibar Martínez, Jose Miguel Miranda
    Nutrients.2026; 18(17): 2889.     CrossRef
  • 3. Case Report: Integrated physiological and performance profiling of an elite athlete with differences of sex development
    Ke Hu, Kaiyun Huang, Xiao Xu, Blair R. Hamilton, Sandra Rozenštoka, Fabio Pigozzi, Samuel Ka-Kin Ling, Ada S. Cheung, Eric Vilain, Yannis P. Pitsiladis
    Frontiers in Sports and Active Living.2026;[Epub]     CrossRef
  • 3,613 View
  • 88 Download
  • Crossref
Prevalence, Clinical Implication, and Cause of Spine Hip Discordance in Elderly Patients with Fragility Hip Fracture
Kyung-Hag Lee, Jung-Wee Park, Sujin Kim, Guen Young Lee, Sung Bin Park, Du-Bin Yang, Yong-Chan Ha
J Bone Metab 2022;29(1):51-57.
Published online February 28, 2022
DOI: https://doi.org/10.11005/jbm.2022.29.1.51
Background
Spine-hip discordance (SHD) increases fracture risk. However, its prevalence and clinical implications have not been investigated in patients with hip fractures. This study determined the prevalence and association of SHD with mortality and investigated the cause of SHD in patients with hip fractures.
Methods
This study included patients admitted for fragility hip fractures between 2011 and 2020. All patients underwent dual energy X-ray absorptiometry and anteroposterior and lateral views of the lumbosacral spine during admission. Data on demographics, diagnosis, American Society of Anesthesiologists score, and mortality were collected. A T-score difference of more than 1.5 between L1-4 and the femur neck was considered discordant, and 3 groups (lumbar low [LL] discordance, no discordance [ND], and femur neck low [FL] discordance) were compared. In the discordance group, lumbar radiographs were reviewed to determine the cause of discordance.
Results
Among 1,220 eligible patients, 130 were excluded due to patient refusal or bilateral hip implantation; therefore, this study included 1,090 patients (271 male and 819 female). The prevalence of LL, ND, and FL was 4.4%, 66.4% and 29.2% in men and 3.9%, 76.1%, and 20.0% women. Mortality was not associated with discordance. The most common causes of discordance were physiological in the LL group and pathological in the FL group for both sexes.
Conclusions
Patients with hip fractures showed lower rates of ND and higher rates of FL compared to the general population. True discordance should be carefully judged for pathological and artifact reasons. The clinical implications of SHD require further investigation.

Citations

Citations to this article as recorded by  Crossref logo
  • 1. Discordance in bone mineral density between lumbar spine and hip in patients with rheumatoid arthritis: A cross-sectional study
    Takeshi Mochizuki, Koichiro Yano, Naoko Koenuma, Katsunori Ikari, Ken Okazaki
    Osteoporosis and Sarcopenia.2026; 12(2): 98.     CrossRef
  • 2. Sex-specific association between lumbar spine-total hip T-score discordance and all-cause mortality: a prospective cohort study from NHANES
    Rıdvan Erten, Mehmet Poyrazer
    Archives of Osteoporosis.2026;[Epub]     CrossRef
  • 3. The observation that older men suffer from hip fracture at DXA T-scores higher than older women and a proposal of a new low BMD category, osteofrailia, for predicting fracture risk in older men
    Yì Xiáng J. Wáng, Ben-Heng Xiao, Jason C. S. Leung, James F. Griffith, Maria Pilar Aparisi Gómez, Alberto Bazzocchi, Davide Diacinti, Wing P. Chan, Ali Guermazi, Timothy C. Y. Kwok
    Skeletal Radiology.2025; 54(5): 925.     CrossRef
  • 4. Assessment of bone turnover markers and DXA parameters to predict bone metastasis progression during zoledronate treatment: a single-center experience
    Stella D’Oronzo, Mauro Cives, Eleonora Lauricella, Stefania Stucci, Antonella Centonza, Marica Gentile, Carmela Ostuni, Camillo Porta
    Clinical and Experimental Medicine.2024;[Epub]     CrossRef
  • 5. Divergent associations of inflammatory markers with bone turnover markers in elderly patients with osteoporotic fractures
    Jian Xu, Yue-qin Guo, Shao-han Guo, Min-zhe Xu, Chong Li, Ya-qin Gong, Ke Lu
    Scientific Reports.2024;[Epub]     CrossRef
  • 6. Different Pattern of T-Score Discordance between Patients with Atypical Femoral Fracture and Femur Neck Fracture
    Byung-Ho Yoon, Jang-Won Park, Chan Woo Lee, Young Do Koh
    Journal of Bone Metabolism.2023; 30(1): 87.     CrossRef
  • 6,552 View
  • 108 Download
  • Crossref
Does Bone Mineral Density Differ between Fan-Beam and Pencil-Beam?: A Meta-Analysis and Systemic Review
Byung-Ho Yoon, Young Do Koh, Jun-Il Yoo, Sujin Kim, Guen Young Lee, Sung Bin Park, Yong-Chan Ha
J Bone Metab 2021;28(1):67-77.
Published online February 28, 2021
DOI: https://doi.org/10.11005/jbm.2021.28.1.67
Background
Dual energy X-ray absorptiometry (DXA) has evolved from pencil-beam (PB) to narrow fan-beam (FB) densitometers. We performed a meta-analysis of the available observational studies to determine how different modes of DXA affect bone mineral density (BMD) measurements.
Methods
A total of 1,233 patients (808 women) from 14 cohort studies were included. We evaluated the differences in BMD according to the DXA mode: PB and FB. Additionally, we evaluated the differences in BMD between the 2 types of FB mode: FB (Prodigy) and the most recent FB (iDXA). Pairwise meta-analysis was performed, and weighted mean differences (WMD) were calculated for (total lumbar, total hip, and total body).
Results
No significant difference was observed in total lumbar (pooled WMD, -0.013; P=0.152) and total hip BMD (pooled WMD, -0.01; P=0.889), between PB and FB. However, total body BMD was significantly lower in the PB compared to the FB group (pooled WMD, -0.014; P=0.024). No significant difference was observed in lumbar BMD (pooled WMD, -0.006; P=0.567), total hip (pooled WMD, -0.002; P=0.821), and total body (pooled WMD, 0.015; P=0.109), between Prodigy and iDXA.
Conclusions
The results of this study warrant the recommendation that correction equations should not be used when comparing BMD from different modes. Further research is still needed to highlight the ways in which differences between DXA systems can be minimized.

Citations

Citations to this article as recorded by  Crossref logo
  • 1. Cross-Calibration Study of The Stratos And Hologic QDR 4500A Dual-Energy X-Ray Absorptiometers to Assess Bone Mineral Density And Body Composition
    Laurent Maïmoun, Sandrine Alonso, Krishna Kunal Mahadea, Vincent Boudousq, Thibault Mura, Denis Mariano-Goulart
    Journal of Clinical Densitometry.2023; 26(4): 101434.     CrossRef
  • 9,343 View
  • 110 Download
  • Crossref